NuZar U Light water system set, produces ultrapure water (Type I water) and RO water from tap water directly. Quality of ultrapure water meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.
The NuZar water purification systems are compact and powerful. They are specifically designed for laboratories that demand a balance between cost of ownership and exceeding their purified water requirements.
The system is CE/RoHS certified. RephiLe provides full documents support to meet user's GMP, GSP, GAP, GCP, GLP requirements.
General Specifications |
|
Operating voltage |
24 VDC |
Input voltage |
100 – 240 VAC |
Main system power |
< 190 W |
Feed Water Requirements |
|
Feed water |
Tap water |
Feed water conductivity / TDS |
< 2000 μS/cm/ < 1000 ppm |
Operating temperature |
5 – 35 °C |
Feed water pressure |
1 – 6 bar psi (15 – 90) |
Product Water Quality |
|
Ultrapure water dispensing rate |
Up to 2 L/min |
Ultrapure water resistivity (@25°C) |
18.2 MΩ·cm |
Ultrapure water TOC |
< 5 ppb* |
Particles in ultrapure water |
No Particles with size > 0.22 μm (with a 0.2 μm final filter or RephiBio filter) |
Microorganisms in ultrapure water |
< 0.001 cfu/ml (with a 0.2 μm final filter or RephiBio filter) |
Pyrogens (endotoxins) in ultrapure water |
< 0.001 Eu/ml (with a RephiBio filter) |
RNase in ultrapure water |
< 0.5 pg/ml (with a RephiBio filter) |
DNase in ultrapure water |
< 10 pg/ml (with a RephiBio filter) |
RO production rate |
24 L/hr |
RO water dispensing rate |
Up to 2 L/min |
RO water conductivity (@25°C) |
Typically < 20 μS/cm |
RO rejection |
97 to 98% ionic rejection (new RO cartridge) > 99% organic rejection > 99% particulates & bacteria rejection |
Dimensions |
|
Main system dimensions Width × depth × height
w/ 10L tank
|
491 x 422 x 624 mm (19.3 x 16.6 x 24.5 in)
|
Main system dimensions Width × depth × height
w/o 10L tank
|
491 x 310 x 624 mm (19.3 x 12.2 x 24.5 in)
|
Dispenser dimensions
Width × depth × height
|
21 × 29 × 61 cm
(8.3 × 11.4 × 24.0 in)
|
Weight (Net approx) |
21 kg |
Weight (Gross approx) |
30 kg |
*When TOC Level of feed water is < 50 ppb.