NuZar water system set, Q produces ultrapure water (Type I water) from RO, distilled or deionized water. Quality of ultrapure water meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.
The NuZar water purification systems are compact and powerful. They are specifically designed for laboratories that demand a balance between cost of ownership and exceeding their purified water requirements.
The system is CE/RoHS certified. RephiLe provides full documents support to meet user's GMP, GSP, GAP, GCP, GLP requirements..
General Specifications |
|
Operating voltage |
24 VDC |
Input voltage |
100 – 240 VAC |
Main system power |
< 190 W |
Feed Water Requirements |
|
Feed water |
EDI water
RO water
Distilled water
Deionized water
|
Operating temperature |
5 – 35 °C |
Feed water pressure |
0 – 1 bar (0 – 15 psi)* |
Product Water Quality |
|
Ultrapure water dispensing rate |
Up to 2 L/min |
Ultrapure water resistivity (@25°C) |
18.2 MΩ·cm |
Ultrapure water TOC |
< 5 ppb** |
Particles in ultrapure water |
No Particles with size > 0.22 μm (with a 0.2 μm final filter or RephiBio filter) |
Microorganisms in ultrapure water |
< 0.001 cfu/ml (with a 0.2 μm final filter or RephiBio filter) |
Pyrogens (endotoxins) in ultrapure water |
< 0.001 Eu/ml (with a RephiBio filter) |
RNase in ultrapure water |
< 0.5 pg/ml (with a RephiBio filter) |
DNase in ultrapure water |
< 10 pg/ml (with a RephiBio filter) |
Dimensions |
|
Main system dimensions Width × depth × height
|
414 x 310 x 624 mm (16.3 x 12.2 x 24.5 in)
|
Weight (Net approx) |
18 kg |
Weight (Gross approx) |
27 kg |
*It’s recommended to use a pressure regulator if feed water pressure is above 15 psi (1 bar)
**When TOC Level of feed water is < 50 ppb.